C-5. U.S. Army Limited or Special Calibration Label (DA Label 163, Figure C-2)
NOTE
Figure C-2 is depicted larger than the actual DA Label 163.
a. Uses:
(1) Verifies that the instrument received a limited calibration, in that certain
parameters or ranges (identified on the label) required repair and could not be
calibrated, or certain parameters and/or ranges were not calibrated, or certain
parameters and/or ranges were calibrated to less stringent tolerances than those
specified in the approved calibration procedures.
(2) Verifies that the instrument received a special calibration in that certain
parameters or ranges (identified on the label) were calibrated to more stringent
tolerances than those specified in the approved calibration procedure and the label
serving as a calibration report. Written request is required from owner of the item.
(3) Indicates that with normal care in handling and use during the calibration
interval (indicated by the date calibrated and the date calibration is void) the special
calibration of the instrument will have an acceptable probability of remaining within the
special calibration tolerances.
b. Preparation. The individual who accomplishes the limited or special calibration
of an instrument will complete DA Label 163. Entries in blocks 1 through 6 will be the
same as outlined in paragraph C-4b above for DA Label 80. Entries in blocks 7 and 8 will
be as follows:
(1) DA Label 163 for limited calibration (fig. C-2).
(a) Block 7 - Line out (obliterate) the words "Special Calibration." Do not line
out (obliterate) the words "Limited Calibration - Do Not Use."
(b) Block 7a - Parameter/value. Enter the parameters, ranges or values that
were not calibrated, are inoperative, or are out of tolerance. Ranges or values not listed
on DA Label 163 must be calibrated to tolerances specified in the approved calibration
procedures.
(c) Block 7b - Tolerance/Uncertain. When parameters, ranges or values have
not been calibrated, enter "NOT CALIBR."  For parameters, ranges, or values that are
out of tolerance, enter out-of-tolerance values or percentage of accuracy's.  For
parameters, ranges, or values that are not operative, enter "INOPERATIVE."
C-3