TB 750-25
USES AND PREPARATION OF DA CALIBRATION LABELS AND FORMS
C-1. Purpose. This appendix provides instructions and procedures for the use and
preparation of the U.S. Army Calibrated Instrument (DA Label 80), the U.S. Army
Limited or Special Calibration (DA Label 163), and the U.S. Army Calibration System
Rejected Instrument (DA Form 2417).
C-2. Applicability. This appendix is applicable to instrument owners/users, DS/GS/
AVIM units, TSAs, and the USAPSL.
C-3. General. Elements of all Army activities that provide TMDE support will use the
DA Label 80, DA Label 163, and DA Form 2417. Overprinting of DA Labels 80 and 163 is
authorized in accordance with paragraph C-7 below.
NOTE
DA Label 80.
a. Uses:
(1) Verifies that the instrument was calibrated to the specifications listed in an
approved calibration procedure or specifications stated on the calibration report
identified in block 4.
(2) Indicates that with normal care in handling and use during the calibration
interval (indicated by the date calibrated and the date calibration is void), the operating
characteristics of the instrument will have an acceptable probability of remaining within
the specified tolerances listed in the approved calibration procedure for the instrument,
or uncertainties given on the calibration report identified in block 4.
b. Preparation. The individual who accomplishes calibration of an instrument will
complete a DA Label 80 as follows:
(1) Block 1 - Date Calibrated.
Enter the calendar date that calibration was
completed.
(2) Block 2 - Calibrated. The activity which performs the calibration service will
be identified by entering the activity's UIC.
NOTE
A rubber stamp may be used to identify the calibration
activity (block 2) and the calibrator (block 4).
|