TB 750-25
(3) Block 3 - Calibration Void. Enter the calendar date on which the calibration
will expire. This date must be based on the interval in TB 43-180, paragraph 4-k, or the
interval requested on DA Form 3758-R ,or by memorandum for TSCs. When a
calibration interval other than calendar days is used for an instrument, enter the
appropriate interval, i.e., 1,000 passes, 2,000 hours, etc.
(4) Block 4 - Name/Report No. The individual who accomplishes calibration will
be identified, or when a calibration test report is required and provided for use with the
instrument, enter the test report number. When two or more individuals accomplish
calibration, only one individual will be identified.
(5) Block 5 - Identification No. Enter the serial number assigned to the
instrument by the manufacturer or the JETDS number listed in TB 43-180. Use the
JETDS serial number when both are provided. If a serial number has not been assigned
by the manufacturer enter a locally assigned number (owner/user generated through the
property book officer) in block 5. Enter the serial number exactly as shown on the
instrument data plate or stamping to include leading zeros, slashes, dashes, etc. Use
the last 10 characters of the serial number.
(6) Block 6 - Owner. Enter the instrument owner's UIC.
NOTE
Blocks 1 through 6 must contain the above data on each DA
Label 80 of calibrated TMDE. Additional data is entered as
deemed necessary, but it should not obscure or contradict the
required data.