TB 750-25
3-5. Procedures
a. X-ray high voltage instruments will be calibrated and repaired by the USAPSL.
To facilitate this task a float of Dynalyzer II and III high voltage instruments has been
established at the USAPSL. The float is used to ensure the AMEDD units have
calibrated instruments continuously available. USATA controls the float and recall
system by shipping a calibrated instrument to AMEDD units.
To assure the
maintenance of an effective operational float system, responsible AMEDD units must
ship their instruments to USATA within 24 hours after receipt of calibrated instruments
from USATA.
(1) Dynalyzer II and III high voltage instruments are handled on an exchange
basis.
(2) For all other makes or models, the original unit will be returned to the
customer upon completion of calibration by USAPSL.
Laboratory. Other TMDE associated with medical x ray equipment is supported in
-
accordance with TB 43-180.
c. The Surgeon General periodically reviews all calibration and repair
requirements and the effect of calibration in the AMEDD health care program.
Section IV. COMMERCIAL CALIBRATION AND REPAIR
elements that receive TMDE services from commercial sources.
3-7. Procedures
a. When a contractor is performing calibration and repair support services which
are the same as the mission of USATA support organizations, the contract will be
written to contain the same regulatory controls as those which apply to the Army.
Calibration and repair support services contracts to be performed in government
facilities with government standards will specify adherence to AR 750-43 and this
pamphlet. Contracts for calibration and repair support which provide for performance
at the contractor's facility using their own standards must specify, as a minimum,
adherence to ANSNI Z540.1-94.
b. Commercial contractors providing calibration support for Army TMDE must
maintain a calibration system which conforms to ANSNI Z540.1-94. Army activities
electing to use calibration services from a commercial source (contractor) are
responsible for obtaining an audit of the contractor's calibration system. A statement
from the contractor attesting to compliance with the standard is not acceptable.
Documentation must be kept on file attesting to the date and by whom the audit was
performed.
|