NOTE
An instrument requiring repair will continue to receive a
cyclic limited calibration until repair is accomplished if it has
functions that are operable. Ensure that repaired TMDE is
calibrated after repairs are accomplished.
The parts
required to accomplish repair will be requisitioned and the
requisition/ document number or work order number entered
along the bottom line of blocks 7a and 7b before the item is
returned to the owner/user.
(d) Block 8 - Authentication.  The supervisor/commander (or designated
alternate) of the instrument owner signs this block indicating the instrument limitations
are known.
(2) DA Label 163 for special calibration (fig C-3).  When an instrument is
calibrated to more stringent specifications other than those listed in the approved
calibration procedure, to a specific value(s), or IAW a locally developed procedure,
entries will be as follows:
(a) Block 7 - Line out (obliterate) the words "Limited Use - Do Not Use." Do
not line out (obliterate) the words "Special Calibration - Corrections Are."
(b) Block 7a - Parameter/Value. Enter the parameters, ranges, or values that
were calibrated to accuracy's or tolerances other than those specified in the approved
calibration procedure.
(c) Block 7b - Tolerance/Uncertain. Enter the tolerance or uncertain that
applies to the parameters or value that was corrected. This entry must be on line with
the parameter/value.
NOTE
When all parameters, ranges, values, and uncertainties
cannot be entered in blocks 7a and 7b, a correction chart or
test report must be provided. The number of the test report
must be entered in block 4 and the words "See Report" must
be entered in blocks 7a and 7b.
(d) Block 8 - Authentication. The immediate supervisor (or a designated
alternate) of the technician performing a calibration will sign his/her name in this block,
attesting that the data entered thereon is an accurate statement of the calibration
results, and the item is released for use. This label will not be affixed to the instrument
without an authenticating signature.